dc.description.abstractThe calibration of the Marconi Moisture meter type TF 933C was done for pepper (capsicum sp.). A kilogram of fresh pepper was blanched and milled into a paste. The paste was subsequently dehydrated using a hot air cabinet dryer. During the drying down process, three samples were taken at a time, at intervals of 4-0 min. In all eight sample batches were taken and the moisture content of the sample batch was determined by the air ventilated oven. After each determination the respective reading on the moisture meter was taken at a room temperature of 28°C. A calibration curve of equation Y = 6.3x was obtained to serve as a reference index for the moisture content of pepper. A relationship was establish by means of a graph and a table expressing readings on the moisture meter corresponding to moisture levels as determined by the air-ventilated oven methoden_US
dc.language.isoenen_US
dc.publisherCouncil for Scientific and Industrial Research (CSIR),Food Research Institute, Ghanaen_US
dc.subjectPepperen_US
dc.subjectCapsicum annuumen_US
dc.subjectCalibrationen_US
dc.subjectMarconi moisture meteren_US
dc.titleCalibration of marconi moisture meter type of 933C for pepper (Capsicum annuum L.)en_US
dc.typeTechnical Reporten_US
Appears in Collections:Food Research Institute

Files in This Item:
File Description SizeFormat 
Calibration_Marconi_Moisture_Meter_Type_for_Pepper.pdf6.1 MBAdobe PDFThumbnail
View/Open


Items in CSIRSpace are protected by copyright, with all rights reserved, unless otherwise indicated.